FDA Medical devices recall ·
Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model recall
- Recall number
- Z-1302-2026
- Date
- (initiated December 11, 2025)
- Company / brand
- Fujirebio Diagnostics, Inc.
201 Great Valley Pkwy, Malvern, PA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1569 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Why was it recalled?
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Model/Catalog Number: 81550; UDI-DI: 05414470815500; Lot Numbers: D4C6028U, D4C6029U, D4C6051U, D4C6052U;
Distribution
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
States: Arizona California Florida Indiana Kansas Kentucky Massachusetts Maryland Minnesota New Jersey Pennsylvania Texas Washington
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1302-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.