FDA Medical devices recall ·

TBS Osteo Product Name: TBS iNsight Model recall

Recall number
Z-1250-2025
Date
(initiated February 3, 2025)
Company / brand
Medimaps Group
Fongit, Plan-les-Ouates, Switzerland
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
2036 units (724 US, 1312 OUS)
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: TBS Osteo Product Name: TBS iNsight Model/Catalog Number: TBS iNsight v3.x / pn: OS-TBSi Software Version: V3.x: (1) V3.1.2, (2) V3.1.1, (3) V3.1.0, (4)V3.0.0 Product Description: TBS iNsight is a class II Medical Device software that is installed on bone densitometers for analysis of bone microarchitecture and bone health management Component: No

Why was it recalled?

Potential variability in calculations from fast array scans compared to array scans when operating on Hologic Horizon machines.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: (1) B214OSTBSI3120, (2) B214OSTBSI3110, (3) B214OSTBSIV310, (4) B214OSTBSIV300.

Distribution

Domestic: AL, AZ, CA, CO, DE, FL, IL, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, PA, SC, TX, UT, WA. International: ALGERIA, ANDORRA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, SOUTH KOREA, LEBANON, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, NEPAL, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY

States: Alabama Arizona California Colorado Delaware Florida Illinois Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi North Carolina New Mexico New York Ohio Pennsylvania Puerto Rico South Carolina Texas Utah Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1250-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.