FDA Medical devices recall ·

Chest Drainage Units and Accessories: Product Code Description 1… recall

Recall number
Z-1203-2026
Date
(initiated December 26, 2025)
Company / brand
Cardinal Health 200, LLC
3651 Birchwood Dr, Waukegan, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1,169,726 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL

Why was it recalled?

The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All units affected by IFU update. UDI-DI: 1. 8888571562 Cardinal Health Sentinel Seal CDU 50192253003091 (CS) 10192253003093 (EA) 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 50192253003084 (CS) 10192253003086 (EA) 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 50192253003077 (CS) 10192253003079 (EA) 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 50192253003039 (CS) 10192253003031 (EA) 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 50192253003015 (CS) 10192253003017 (EA) 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 50192253003060 (CS) 10192253003062 (EA) 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 50192253003022 (CS) 10192253003024 (EA) 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 50192253002643 (CS) 10192253002645 (EA) 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 50192253002308 (CS) 10192253002300 (CS) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 50192253002292 (CS) 10192253002294 (EA) 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 50192253002650 (CS) 10192253002652 (EA) 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL 50192253002667 (CS) 10192253002669 (EA)

Distribution

Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1203-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.