FDA Medical devices recall ·
Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001… recall
- Recall number
- Z-1202-2025
- Date
- (initiated January 27, 2025)
- Company / brand
- Dexcom, Inc.
6340 Sequence Dr, San Diego, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- N/A
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001, STR-OR-001
Why was it recalled?
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
All Serial Numbers/UDI: 00386270000583
Distribution
Worldwide - US Nationwide distribution in the states of OH, TX, PA, MN, SC, NH, FL, CA, NJ, PR, GA and the countries of Austria, Belgium, Canada, Switzerland, Germany, Spain, France, Ireland, Japan, Netherlands, Slovenia, United Kingdom, Australia, Czech Republic, Denmark, Finland, France, Hong Kong, Israel, Italy, Korea, Kuwait, New Zealand, Norway, Oman, Poland, Qatar, Saudi Arabia, Singapore, Slovakia, South Africa, Sweden, Turkey, United Arab Emirates.
States: California Florida Georgia Minnesota New Hampshire New Jersey Ohio Pennsylvania Puerto Rico South Carolina Texas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1202-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.