FDA Medical devices recall ·

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode… recall

Recall number
Z-1197-2025
Date
(initiated December 16, 2024)
Company / brand
Augustine Temperature Management, LLC
15305 Minnetonka Blvd, Minnetonka, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
4 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

Why was it recalled?

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297

Distribution

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

States: Alabama Nevada Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1197-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.