FDA Medical devices recall ·
stryker Precision Thin recall
- Recall number
- Z-1138-2025
- Date
- (initiated January 8, 2025)
- Company / brand
- Stryker Corporation
1941 Stryker Way, Portage, MI, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 622 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
stryker Precision Thin, Catalog Number REF 2296-003-108; Micro Oscillating and Sagittal Blades
Why was it recalled?
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
GTIN 04546540046482, Lot Number 22298017
Distribution
Worldwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.