FDA Medical devices recall ·

ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US)… recall

Recall number
Z-1118-2026
Date
(initiated December 15, 2025)
Company / brand
ICU Medical, Inc.
600 N Field Dr, Lake Forest, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)

Why was it recalled?

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

400020401 (US), Software Versions: 1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions: 1.1.0

Distribution

Worldwide distribution - US Nationwide and the country of Philippines.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1118-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.