FDA Medical devices recall ·
Tego Connector recall
- Recall number
- Z-1111-2026
- Date
- (initiated December 19, 2025)
- Company / brand
- ICU Medical, Inc.
951 Calle Amanecer, San Clemente, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 14,684,607 items
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
Why was it recalled?
Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 00840619026059, 00840619006785. REF/Lot(Expiration): 011-D1000/14015455(5/1/2029), 14044313(6/1/2029), 14228823(12/1/2029), 14274294(1/1/2030); 01C-D1000/14171414(10/1/2029), 14304670(2/1/2030); 055-D1000/13772668(9/1/2028), 13778926(9/1/2028), 13791786(10/1/2028), 13794986(10/1/2028), 13797963(10/1/2028), 13801102(10/1/2028), 13806764(10/1/2028), 13812991(11/1/2028), 13822705(11/1/2028), 13828556(11/1/2028), 13833879(11/1/2028), 13838937(12/1/2028), 13867515(1/1/2029), 13875052(1/1/2029), 13876502(1/1/2029), 13876503(1/1/2029), 13882254(1/1/2029), 13882256(1/1/2029), 13887995(1/1/2029), 13887996(1/1/2029), 13894349(2/1/2029), 13894350(2/1/2029), 13903971(2/1/2029), 13903972(2/1/2029), 14045054(6/1/2029), 14045055(6/1/2029), 14049379(6/1/2029), 14049404(6/1/2029), 14063784(6/1/2029), 14063785(6/1/2029), 14071890(7/1/2029), 14071891(7/1/2029), 14071892(7/1/2029), 14087320(7/1/2029), 14090798(7/1/2029), 14110500(8/1/2029), 14131560(8/1/2029), 14144999(9/1/2029), 14145000(9/1/2029), 14150873(9/1/2029), 14150874(9/1/2029), 14150875(9/1/2029), 14170224(10/1/2029), 14170226(10/1/2029), 14179702(10/1/2029), 14179704(10/1/2029), 14179706(10/1/2029), 14208522(11/1/2029), 14208523(11/1/2029), 14212866(11/1/2029), 14212870(11/1/2029), 14216154(11/1/2029), 14216155(11/1/2029), 14224623(11/1/2029), 14248560(12/1/2029), 14248561(12/1/2029), 14274291(1/1/2030), 1427429291/1/2030), 14278257(1/1/2030), 14278258(1/1/2030), 14278259(1/1/2030), 14278260(1/1/2030), 14300970(2/1/2030), 143
Distribution
Worldwide - US Nationwide distribution including in the states of FL, AZ, RI, TN, KS, NJ, NY, VA, SC, GA, CA, MS, TX, OH, NC, MA, WA, IL, PA, NV, NH, LA, MI, NE, ME, MN, NM, OK, MD, IA, OR, SD, WI, MO, AL, WY, IN, KY, DC, UT, AR, WV, AK, ND, CO and the countries of BE, DE, CO, ES, AU, AR, SA, SI, IT, LK, MY, AT, FI, GB, TW, MX, NL, IE, CA, GR, ZA, SE, SK, NZ, LT, CH, HR, FR, JO, PE, SG, BR, PL, RO, HU, CZ, NO, AE, EE, CN, MA, HK.
States: Alaska Alabama Arkansas Arizona California Colorado District of Columbia Delaware Florida Georgia Iowa Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin West Virginia Wyoming
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1111-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.