FDA Medical devices recall ·
RayStation Product Name: RayStation/RayPlan Model recall
- Recall number
- Z-1110-2026
- Date
- (initiated November 28, 2025)
- Company / brand
- Raysearch Laboratories Ab
Eugeniavagen 18c, Stockholm, Sweden - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 0
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Why was it recalled?
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Lot Code: UDI: 0735000201082220250423, 0735000201094520250702 GTIN: 07350002010822, 07350002010945 Serial Numbers: 17.0.0.1270, 17.0.1.113 Software Revision: RayStation v2025, RayStation v2025 SP1 Expiration Date: 2030-10-22
Distribution
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1110-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.