FDA Medical devices recall ·

Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP… recall

Recall number
Z-1101-2025
Date
(initiated January 10, 2025)
Company / brand
Nobel Biocare
Vastra Hamngatan 1, Goteborg, Sweden
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1,558 units (90 units - U.S.)
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Why was it recalled?

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Nobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213

Distribution

U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

States: California Florida Illinois Massachusetts New Jersey New York Ohio Pennsylvania Utah Virginia Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1101-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.