FDA Medical devices recall ·
MOSAIQ Oncology Information System with Particle Therapy License (License… recall
- Recall number
- Z-1078-2026
- Date
- (initiated December 10, 2025)
- Company / brand
- Elekta, Inc.
400 Perimeter Center Ter NE Ste 50, Atlanta, GA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Why was it recalled?
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071
Distribution
US Nationwide distribution in the state of TX.
States: Texas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1078-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.