FDA Medical devices recall ·

MOSAIQ Oncology Information System with Particle Therapy License (License… recall

Recall number
Z-1078-2026
Date
(initiated December 10, 2025)
Company / brand
Elekta, Inc.
400 Perimeter Center Ter NE Ste 50, Atlanta, GA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Why was it recalled?

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071

Distribution

US Nationwide distribution in the state of TX.

States: Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1078-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.