FDA Medical devices recall ·

VITROS Product Name: VITROS XT 7600 Integrated System Model recall

Recall number
Z-1006-2025
Date
(initiated December 19, 2024)
Company / brand
Ortho-Clinical Diagnostics, Inc.
513 Technology Blvd, Rochester, NY, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1,893 units (994 US, 899 OUS)
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified), 6904003 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component

Why was it recalled?

The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: VITROS XT 7600 Integrated System Product Code: 6844461 (New) UDI: 10758750012343 VITROS XT 7600 System (Certified) Product Code: 6272222 UDI: 10758750012343 Lot Code: VITROS XT 7600 Integrated System Product Code: 6904003 (China) UDI: 10758750035809 Range of installed serial numbers:76000001-76002199 Manufacture date range: July 2018 to present Distribution date range; July 2018 to present Total systems currently installed in the US: 994 Total systems currently installed outside the US: 899 Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems t

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1006-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.