FDA Medical devices recall ·

StatStrip Glucose Hospital Meter System recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-0681-2025
Date
(initiated November 20, 2024)
Company / brand
Nova Biomedical Corporation
200 Prospect St, Waltham, MA, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
2,365 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.

Why was it recalled?

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model No 63685. UDI-DI (01)10385480636858. Software Versions: v0.0.13.10 to v0.0.13.44

Distribution

Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

States: Puerto Rico

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-0681-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.