Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20… recall

Recall number
Z-0597-2026
Date
(initiated October 27, 2025)
Company / brand
Philips North America Llc
222 Jacobs St, Cambridge, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
110
Initiated by
FDA Mandated

Read the official FDA notice ↗

Product

Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM

Why was it recalled?

Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model Number 722234 Commercial Name: Azurion 7 M20 How to identify: Philips Reference C&R 2025-IGT-BST-016