FDA Medical devices recall ·
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL recall
- Recall number
- Z-0554-2025
- Date
- (initiated October 9, 2024)
- Company / brand
- Shimadzu Corp. Analytical & Measuring Instruments Div.
1 Chome 1, Kyoto, Japan - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 48
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Why was it recalled?
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.
Distribution
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
States: Delaware Pennsylvania
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0554-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.