FDA Medical devices recall ·

Z-800 Infusion System recall

Recall number
Z-0506-2025
Date
(initiated October 18, 2024)
Company / brand
Zyno Medical LLC
177 Pine St, Natick, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
34,994 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.

Why was it recalled?

Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200.

Distribution

This product is distributed only in the United States.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0506-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.