FDA Medical devices recall ·

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name:… recall

Recall number
Z-0498-2025
Date
(initiated September 18, 2024)
Company / brand
Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1749 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Why was it recalled?

To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model: GIF-1TH190, UDI-DI: 04953170343360, Serial #: Less than 2446103.

Distribution

Domestic: US Nationwide Distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0498-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.