FDA Medical devices recall ·

Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve… recall

Recall number
Z-0491-2025
Date
(initiated October 17, 2024)
Company / brand
Abbott Medical
177 E County Road B, Saint Paul, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1 US, 9 OUS
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile

Why was it recalled?

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI (GTIN): 05415067045768, Serial Numbers: 20078589, 20357219, 20357226, 20357685, 20358319, 20358343, 20358351, 20358363, 20358368, & 20358370.

Distribution

New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

States: New York

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0491-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.