FDA Medical devices recall ·

SimplyGo Mini, Standard Battery Kit recall

Recall number
Z-0481-2026
Date
(initiated October 7, 2025)
Company / brand
Philips Respironics, Inc.
1001 Murry Ridge Ln, Murrysville, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
56 units (55 US, 1 OUS)
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600

Why was it recalled?

Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;

Distribution

US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.

States: Georgia Idaho Kentucky New Jersey New York

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0481-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.