FDA Medical devices recall ·

Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer… recall

Recall number
Z-0452-2026
Date
(initiated October 8, 2025)
Company / brand
Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1528 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System

Why was it recalled?

The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model/Catalog Number: SPL-T; UDI-PI: 00855279005023; All serial numbers;

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0452-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.