FDA Medical devices recall ·

SAFE-T-FILL Micro Capillary Blood Collection… recall

Recall number
Z-0372-2026
Date
(initiated May 5, 2025)
Company / brand
Anatomy Supply Partners
7800 Third Flag Pkwy, Austell, GA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
4,272,858
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910

Why was it recalled?

All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.

States: Alaska Alabama Arizona California Colorado Connecticut Florida Georgia Iowa Idaho Illinois Indiana Kentucky Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi North Carolina Nebraska New Jersey New York Ohio Pennsylvania South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0372-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.