FDA Medical devices recall ·
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT… recall
- Recall number
- Z-0356-2027
- Date
- (initiated October 2, 2026)
- Company / brand
- Corin Medical, Ltd.
Unit 1-57-8, Cirencester, United Kingdom - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 22 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
Why was it recalled?
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
States: California Georgia Illinois Michigan North Carolina Rhode Island Texas Utah
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0356-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.