FDA Medical devices recall ·

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT… recall

Recall number
Z-0356-2027
Date
(initiated October 2, 2026)
Company / brand
Corin Medical, Ltd.
Unit 1-57-8, Cirencester, United Kingdom
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
22 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

Why was it recalled?

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.

Distribution

Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.

States: California Georgia Illinois Michigan North Carolina Rhode Island Texas Utah

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0356-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.