FDA Medical devices recall ·

CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 recall

Recall number
Z-0325-2027
Date
(initiated September 18, 2026)
Company / brand
Artivion, Inc
1655 Roberts Blvd Nw, Kennesaw, GA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010

Why was it recalled?

A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Serial/Donor Number: 12617554/177871, UDI-DI: 00877234000478, Shipped: 22May2024, Implanted: 05June2024

Distribution

US Nationwide distribution in the state of IL.

States: Illinois

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0325-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.