FDA Medical devices recall ·

Hillrom Welch Allyn XScribe Cardiac Stress Testing System recall

Recall number
Z-0325-2025
Date
(initiated October 11, 2024)
Company / brand
Baxter Healthcare Corporation
1 Baxter Pkwy, Deerfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
766 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer

Why was it recalled?

There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Software Versions: Software Version V6.0.0 V6.3.2, Release Date 02-Jul-2015 Units: 1) UDI/DI 00732094259032, Serial Numbers: 115340269178, 117060553883, 117060553884, 117060553885, 119230003169; 2) UDI/DI 00732094259070, Serial Numbers: 117130559202, 117130559203, 117130559204; 3) UDI/DI 00732094259100, Serial Numbers: 116270497046, 117190567440; 4) UDI/DI 00732094259148, Serial Numbers: 117350583961, 119060000843, 122090000529; 5) UDI/DI 00732094259179, Serial Numbers: 116240458434, 116500543985, 120480000930; 6) UDI/DI 00732094259230, Serial Numbers: 118160628177; 7) UDI/DI 00732094259261, Serial Numbers: 117470596306; 8) UDI/DI 00732094259285, Serial Numbers: 118100618358, 116110348774, 116490541753, 117430591185, 117510601740; 9) UDI/DI 00732094259292, Serial Numbers: 116460537153, 118490000226; 10) UDI/DI 00732094259308, Serial Numbers: 116120365527, 119410001349, 119410001350, 119410001351, 119410001352, 119410001353, 119410001354, 119410001355, 119410001356, 119410001357, 119410001358, 119410001359, 119410001360, 119410001361, 119410001362, 119410001363, 119410001364, 119410001365, 119410001366, 119410001367, 119410001368, 119410001369, 119410001370; 11) UDI/DI 00732094259315, Serial Numbers: 117130559201; 12) UDI/DI 00732094259322, Serial Numbers: 117350583998, 117500601523; 13) UDI/DI 00732094259346, Serial Numbers: 116110347146, 116110347144; 14) UDI/DI 00732094259445, Serial Numbers: 116430533831, 116430533832; 15) UDI/DI 00732094259506, Serial Numbers: 11638052713

Distribution

Worldwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0325-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.