Hemochron Jr., Whole Blood Microcoagulation System… recall

Recall number
Z-0317-2027
Date
(initiated September 14, 2026)
Company / brand
Accriva Diagnostics, Inc.
6260 Sequence Dr, San Diego, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
28710
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103

Why was it recalled?

Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027)

Distribution

Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.

States: Alaska Illinois Massachusetts Missouri New Jersey New York Texas Virginia West Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0317-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.