FDA Medical devices recall ·
Orthoscan VERSA Accessory Cart recall
- Recall number
- Z-0286-2027
- Date
- (initiated September 3, 2026)
- Company / brand
- Ziehm-Orthoscan Inc
14555 N 82nd St, Scottsdale, AZ, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 55 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020
Why was it recalled?
The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI/ Mini C-Arm serial number (6Axxxx)/Accessory cart serial Number: (VCXXX): 1 6A0110 /VC0110 2 6A0111 /VC0111 3 6A0112 /VC0112 4 6A0115 /VC0113 5 6A0114 /VC0114 6 6A0117 /VC0115 7 6A0116 /VC0116 8 6A0113 /VC0117 9 6A0118 /VC0118 10 6A0119/ VC0119, VC0120; 11 6A0120 /VC0121; 12 6A0121 /VC0122; 13 6A0123 /VC0123; 14 6A0122 /VC0124; 15 6A0126 /VC0125; 16 6A0133 /VC0134; 17 6A0124 /VC0126; 18 6A0128 /VC0128; 19 6A0127 /VC0129; 20 6A0129 /VC0130; 21 6A0125 /VC0127; 22 6A0131 /VC0132; 23 6A0134 /VC0133; 24 6A0142 /VC0144; 25 6A0136 /VC0136; 26 6A0132 /VC0137; 27 6A0137 /VC0138; 28 6A0138 /VC0139; 29 6A0139 /VC0140; 30 6A0140 /VC0141; 31 6A0141 /VC0142; 32 6A0146 /VC0147; 33 6A0143 /VC0143; 34 6A0144 /VC0145; 35 6A0152 /VC0153; 36 6A0147 /VC0148; 37 6A0148 /VC0149; 38 6A0149 /VC0150; 39 6A0150 /VC0151; 40 6A0151 / VC0152; 41 6A0163 / VC0165; 42 6A0130 / VC0131; 43 6A0155 / VC0156; 44 6A0157 / VC0157; 45 6A0156/ VC0158; 46 6A0158 / VC0159, VC0160; 47 6A0159/ VC0161; 48 6A0160/ VC0162; 49 6A0161 / VC0163; 50 6A0162 VC0164; 51 6A0153 VC0154; 52 6A0164 VC0166; 53 6A0165 VC0167
Distribution
Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.
States: Arizona California Colorado Connecticut Florida Georgia Illinois Indiana Kansas Massachusetts Maryland Michigan Minnesota North Carolina Nebraska Nevada New York Ohio Oregon South Carolina Texas Utah Washington
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0286-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.