FDA Medical devices recall ·

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D… recall

Recall number
Z-0284-2025
Date
(initiated September 17, 2024)
Company / brand
Canon Medical System, USA, INC.
2441 Michelle Dr, Tustin, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
27
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Why was it recalled?

When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029

Distribution

US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA

States: Alabama Arkansas Arizona California Colorado Iowa Illinois Indiana Kansas Louisiana Maryland Minnesota Mississippi North Carolina Nevada Tennessee Virginia Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0284-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.