FDA Medical devices recall ·
Vial2Bag Advanced¿ Product Name: Vial2Bag Advanced¿ 20mm Admixture Device… recall
- Recall number
- Z-0283-2025
- Date
- (initiated September 24, 2024)
- Company / brand
- West Pharmaceutical Services AZ, Inc.
7400 E Tierra Buena Ln Ste 101, Scottsdale, AZ, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 129600
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: Vial2Bag Advanced¿ Product Name: Vial2Bag Advanced¿ 20mm Admixture Device Model/Catalog Number: 6070030 Software Version: Not applicable Product Description: The Vial2Bag Advanced 20mm Admixture Device is a single use, fluid transfer device that allows for the reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution, through the IV bag administration port. The device consists of the body, Protector, IV Port, and an integrated vial adapter. The device is provided as a sterile, non-pyrogenic product. The device is intended to be used with standard drug vials with a seal diameter of 20mm and an elastomeric stopper. The Vial2Bag Advanced 20mm Admixture Device is designed to work with a standard 50, 100, or 250mL IV bag and an external IV administration set. The device does not contain any medicinal substances and there are no additional accessories needed or provided with the Vial2Bag Advanced 20mm Admixture Device for the device to meet its intended purpose. Component: No
Why was it recalled?
Due to packaging damage from heat exposure during shipping/transportation that may result in a breach in the sterile barrier.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Lot Code: Recall Product 1: Product Code 6080030, Lot# K188, UDI (01)10850022888011(17)270430(10)K188\F(90)36098150\F(91)6070030 Recall Product 2: Product Code 6080030, Lot# K194, UDI (01)10850022888011(17)270430(10)K194\F(90)36098150\F(91)6070030
Distribution
A U.S. Distributor, Progressive Medical Inc. (PMI) currently supplies the Vial2Bag Advanced 20mm Admixture Device to hospitals in the U.S. (end user). PMI distributed the impacted Vial2Bag Advanced 20mm Admixture Device lots to the end users. U.S.: Distributor located in MO U.S. Customers: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MD, MI, MN, MO, MS, NC, NE, NM, NV, NY, OH, OK, OR, PA,SC, TN, TX, UT, WA, and WI.
States: Alabama Arkansas Arizona California Colorado Connecticut Florida Georgia Hawaii Illinois Indiana Kentucky Louisiana Maryland Michigan Minnesota Missouri Mississippi North Carolina Nebraska New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania South Carolina Tennessee Texas Utah Washington Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0283-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.