FDA Medical devices recall ·

Xact mole marker for mammography Model recall

Recall number
Z-0268-2027
Date
(initiated September 22, 2026)
Company / brand
Solstice Corporation
85 E St, South Portland, ME, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
736
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Xact mole marker for mammography Model/Catalog Number: M15RM

Why was it recalled?

The radiopaque material may detach from the marker. In some cases, artifact was noted and attributed to the detached material.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

part #: M15RM; UDI-DI: 00854623006068; UPC-A; lot #: M15RM26A;

Distribution

Worldwide - US Nationwide distribution in the states and territory of CA, FL, IL, IN, MN, MO, NY, OH, PA, TX, WV and Puerto Rico. The country of Canada.

States: California Florida Illinois Indiana Minnesota Missouri New York Ohio Pennsylvania Puerto Rico Texas West Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0268-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.