FDA Medical devices recall ·
Abbott Alinity m HPV AMP Kit, used with the Alinity m System… recall
- Recall number
- Z-0233-2025
- Date
- (initiated September 12, 2024)
- Company / brand
- Abbott Molecular, Inc.
1300 E Touhy Ave, Des Plaines, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 13318 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
Why was it recalled?
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
a) REF 09N15-095, UDI/DI 00884999048614; b) REF 09N15-090, UDI/DI 00884999047921; c) REF 09N15-091, UDI/DI 00884999049529; All Lots
Distribution
Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA, IN, IL, LA, MA, MI, MN, MO, NC, NE, NJ, NY, OR, PA, SC, TN, TX, VA, WA, and WI; and foreign (OUS) countries of: AUSTRIA, BELGUIM, BULGARIA, CAMBODIA, CANADA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REP., EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, KAZAKHSTAN, KENYA, LUXEMBOURG, MALAWI, MALAYSIA, MONTENEGRO, MOROCCO, NETHERLANDS, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDIA ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TANZANIA, THAILAND, UNITED KINGDOM, AND VIETNAM.
States: Arizona California Colorado District of Columbia Florida Georgia Illinois Indiana Louisiana Massachusetts Michigan Minnesota Missouri North Carolina Nebraska New Jersey New York Oregon Pennsylvania South Carolina Tennessee Texas Virginia Washington Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0233-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.