FDA Medical devices recall ·
EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (>… recall
- Recall number
- Z-0225-2025
- Date
- (initiated September 13, 2024)
- Company / brand
- Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 209 units
- Hazard tags
- Mislabeling
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. Model/Catalog Number: BD-410X-1055
Why was it recalled?
Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 821925033238 Lot: 408987
Distribution
Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.
States: California Delaware Florida Illinois Indiana New Jersey New York Pennsylvania Tennessee Virginia Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0225-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.