FDA Medical devices recall ·

EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (>… recall

Recall number
Z-0225-2025
Date
(initiated September 13, 2024)
Company / brand
Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
209 units
Hazard tags
Mislabeling
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. Model/Catalog Number: BD-410X-1055

Why was it recalled?

Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 821925033238 Lot: 408987

Distribution

Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.

States: California Delaware Florida Illinois Indiana New Jersey New York Pennsylvania Tennessee Virginia Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0225-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.