FDA Medical devices recall ·
CADD-Solis Li-ion Rechargeable Battery Packs recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-0222-2025
- Date
- (initiated July 16, 2024)
- Company / brand
- Smiths Medical ASD, Inc.
6000 Nathan Ln N, Minneapolis, MN, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 138039 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.
Why was it recalled?
Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
All serial numbers.
Distribution
Worldwide Distribution
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-0222-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.