FDA Medical devices recall ·

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with… recall

Recall number
Z-0213-2026
Date
(initiated September 5, 2025)
Company / brand
Wipro GE Healthcare Private Ltd.
122 Part 1, Bengaluru, India
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
173 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube

Why was it recalled?

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

GTIN 00195278343512, Serial Numbers: STU25210418SA, STU25210417SA, STU25210419SA, STU25250440SA, STU25250439SA, STU25260447SA, STU25210407SA, STU25210406SA, STU24200003SA, STU24200006SA, STU24200005SA, STU24200004SA, STU23240015SA, STU24240019SA, STU24240020SA, STU24240007SA, STU24240022SA, STU24240021SA, STU24330020SA, STU24240014SA, STU24240013SA, STU24330019SA, STU24330022SA, STU24330016SA, STU24240005SA, STU24240008SA, STU24240012SA, STU24330018SA, STU24330021SA, STU24330017SA, STU23160011SA, STU23240007SA, STU24310001SA, STU25210433SA, STU25210432SA, STU25200390SA, STU25200392SA, STU25200391SA, STU25200399SA, STU25200397SA, STU25200395SA, STU25200393SA, STU25200398SA, STU25200396SA, STU25250438SA, STU25270475SA, STU25270456SA, STU25270468SA, STU25270458SA, STU25270466SA, STU25270476SA, STU25270469SA, STU25270473SA, STU25270459SA, STU25270474SA, STU25270467SA, STU25210430SA, STU25210429SA, STU25180343SA, STU25190347SA, STU25190348SA, STU24200007SA, STU25190360SA, STU25240436SA, STU25190361SA, STU25190362SA, STU25190363SA, STU25240437SA, STU25190359SA, STU24510027SA, STU24510032SA, STU24510010SA, STU24510020SA, STU24510017SA, STU24510026SA, STU24510018SA, STU24510024SA, STU24510025SA, STU24510031SA, STU24510011SA, STU24510029SA, STU24510023SA, STU24510030SA, STU24510021SA, STU24510019SA, STU24510022SA, STU25190350SA, STU25200388SA, STU25200389SA, STU25210412SA, STU25190349SA, STU25190381SA, STU25190376SA, STU25190383SA, STU25190374SA, STU25190367SA, STU25190372SA, STU25190

Distribution

Worldwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0213-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.