FDA Medical devices recall ·

Techno-Path Multichem AE Unassayed Level 3 recall

Recall number
Z-0207-2027
Date
(initiated August 14, 2026)
Company / brand
Techno-Path Manufacturing Ltd.
Fort Henry Business Park, Ballina Tipperary, Ireland
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
832 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

Why was it recalled?

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI:(01)05391523442824/-Lot # 65705263

Distribution

US Nationwide distribution in the state of Massachusetts.

States: Massachusetts

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0207-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.