FDA Medical devices recall ·

VACUETTE Product Name: LH Lithium Heparin Separator Model recall

Recall number
Z-0200-2027
Date
(initiated July 29, 2026)
Company / brand
Greiner Bio-One North America, Inc.
4238 Capital Dr, Monroe, NC, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
860450
Hazard tags
Microbial contamination
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)

Why was it recalled?

Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12

Distribution

U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.

States: Colorado Florida Illinois Missouri Nebraska New York Ohio Pennsylvania Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0200-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.