FDA Medical devices recall ·

Medical convenience kits Item number/ Product description AHLD15X LABOR… recall

Recall number
Z-0173-2026
Date
(initiated September 4, 2025)
Company / brand
American Contract Systems Inc
1601 Se Gateway Dr Ste 120, Grimes, IA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
32,433 total
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK

Why was it recalled?

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

AHLD15X UDI-DI 191072224767 Lot 8188711 Bag serial no. 17555039 17555044 17555049 17609166 17609181; AKDV96B UDI-DI 191072224019 Lot 7311311 Bag serial no. 47140875 47141919; EVBH28D UDI-DI 191072239051 Lot 8424611 Bag serial no. 17754395; UPVD01K UDI-DI 191072238993 Lot 8439111 Bag serial no. 17682598 Log 8499211 Bag serial no. 17742543

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States: Iowa Illinois Minnesota South Dakota Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0173-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.