FDA Medical devices recall ·

Williams Cystoscopic Injection Needles, used for cystoscopic injection… recall

Recall number
Z-0163-2027
Date
(initiated September 16, 2026)
Company / brand
Cook Incorporated
750 N Daniels Way, Bloomington, IN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
97 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112

Why was it recalled?

The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X

Distribution

US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.

States: Florida Illinois New Mexico Texas Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0163-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.