MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1 recall
- Recall number
- Z-0159-2027
- Date
- (initiated August 17, 2026)
- Company / brand
- Microgenics Corporation
46500 Kato Rd, Fremont, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1366
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP
Why was it recalled?
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)
Distribution
Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL.
States: California Louisiana Michigan Missouri New York Ohio Oklahoma Oregon South Carolina Texas Utah
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0159-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.