FDA Medical devices recall ·

PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture… recall

Recall number
Z-0153-2027
Date
(initiated September 8, 2026)
Company / brand
Ethicon, LLC
Highway 183 Km 8.3, San Lorenzo, PR, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1764
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H

Why was it recalled?

Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026

Distribution

Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.

States: Alabama Arizona California Colorado Florida Idaho Illinois Indiana Louisiana Massachusetts Maine Minnesota Missouri North Carolina Nebraska New Hampshire New York Ohio Oklahoma Pennsylvania South Carolina Tennessee Texas Utah Wisconsin Wyoming

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0153-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.