FDA Medical devices recall ·
IOMAX Product Name: IOMAX Cortical Module Model recall
- Recall number
- Z-0141-2027
- Date
- (initiated August 18, 2026)
- Company / brand
- Cadwell Industries Inc
909 N Kellogg St, Kennewick, WA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 23
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system
Why was it recalled?
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Lot Code: These devices are identified by the serial numbers listed below: 19029607AA0326007 19029607AA0226013 19029607AA0426028 19029607AA0226010 19029607AA0226007 19029607AA0326026 19029607AA0226015 19029607AA0326020 19029607AA0326018 19029607AA0326021 19029607AA0326022 19029607AA0226009 19029607AA0526022 19029607AA0326017 19029607AA0326027 19029607AA0326016 19029607AA0226016 19029607AA0326028 19029607AA0326024 19029607AA0326025 19029607AA0326023 19029607AA0326031 19029607AA0326006
Distribution
Worldwide - US Nationwide distribution in the state of TN and the countries of Spain, Vietnam, Malaysia, Saudi Arabia, Thailand, Israel, Mexico, Turkey, Oman, Colombia, Brazil, Belgium, Italy, South Africa, and Switzerland.
States: Tennessee
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0141-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.