FDA Medical devices recall ·
JenaValve Trilogy Delivery Catheter, Size 25MM recall
- Recall number
- Z-0135-2027
- Date
- (initiated July 31, 2026)
- Company / brand
- Jenavalve Technology Inc.
4 Cromwell, Irvine, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 108 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
JenaValve Trilogy Delivery Catheter, Size 25MM REF: TRILOGY-DC-M
Why was it recalled?
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
GTIN - 00850000898141/Lot # DS-040826-005, DS-040926-002, DS-042326-001, DS-042826-007, DS-042926-008, DS-050826-005, DS-051526-001
Distribution
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy.
States: Arizona California Florida Georgia Illinois Massachusetts Maryland Michigan Minnesota Missouri New York Ohio Pennsylvania South Carolina Texas Utah Virginia Washington
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0135-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.