FDA Medical devices recall ·

Intellijoint Navigation System: Workstation G2 Configured recall

Recall number
Z-0119-2027
Date
(initiated August 28, 2026)
Company / brand
Intellijoint Surgical, Inc.
809 Wellington St N Unit 2, Kitchener, N/A, Canada
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
8
Hazard tags
Fire
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249

Why was it recalled?

A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

REF/UDI-DI/MacBook Serial Numbers: 000-0241/00628184002414/C02RP36YG8WP, C02QPCF5G8WN 000-0250/00628184002506/C02RF44TG8WN, C02QW25RG8WL, C02QH355G8WM, C02S22N9G8WP; 000-0249/00628184002490/C02RQ3FUG8WN, C02R4451G8WN

Distribution

Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.

States: California Mississippi New Jersey Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0119-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.