FDA Medical devices recall ·
Monarch Inflation Device 30 atm/bar MAP302 Honor Hemostasis Valve Metal… recall
- Recall number
- Z-0059-2025
- Date
- (initiated August 30, 2024)
- Company / brand
- Merit Medical Systems, Inc.
1600 W Merit Pkwy, South Jordan, UT, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1,120 devices
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Monarch Inflation Device 30 atm/bar MAP302 Honor Hemostasis Valve Metal Insertion Tool Torque Device REF IN2330 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid.
Why was it recalled?
Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Catalog Number: IN2330/B UDI-DI code: 00884450298046 Lot Number: H2369557 H2416390 H2421992 H2426112 H2492843 H2522880 H2540015 H2554505 H2560208 H2579556 H2765640 H2793109 H2806018 H2810777
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Japan, Monaco, Netherlands, Qatar, United Arab Emirates, United Kingdom.
States: Alabama Arkansas Arizona California Colorado Connecticut Florida Georgia Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming