FDA Medical devices recall ·
MPR Product Name:SINGLE USE TOURNIQUET CUFF 30" 1 LINE (10) Model recall
- Recall number
- Z-0055-2027
- Date
- (initiated August 24, 2026)
- Company / brand
- ASP Global, LLC.
7800 Third Flag Pkwy, Austell, GA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1100
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 30" 1 LINE (10) Model/Catalog Number: DT-7030-01Q Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
Why was it recalled?
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Lot Code: Model Number: DT-7030-01Q UDI-DI: UDI was not applicable at time of manufacture. All Lots produced from 01SEP2016 to 01MAR2023: S1101STY, P1015STY, N0330STY, N0730STY
Distribution
US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
States: Alabama Arkansas Arizona California Colorado Connecticut Florida Georgia Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi North Carolina Nebraska New Jersey New Mexico Nevada New York Ohio Pennsylvania South Carolina Tennessee Texas Virginia Washington Wisconsin Wyoming
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0055-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.