FDA Medical devices recall ·

Enterprise Imaging XERO Viewer- software application used for reference… recall

Recall number
Z-0055-2025
Date
(initiated September 6, 2024)
Company / brand
Agfa Healthcare NV
Septestraat 27, Mortsel, Belgium
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
801 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)

Why was it recalled?

Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 05400874000710 Software Versions affected: All versions of Enterprise Imaging XERO Viewer are affected 8.1.2 or higher 8.1.4 or higher 8.2.0.100 or higher 8.2.2.000 or higher 8.3.2.000 or higher 8.4.0.000 or higher Enterprise Imaging XERO Viewer main version streams 8.4.0, 8.4.1, 8.3.2, 8.3.1, 8.3.0, 8.2.2, 8.2.1, 8.2.0, 8.1.5, 8.1.4, 8.1.3, 8.1.2, 8.1.1, 8.1, 8.0.1, 8.0.0, ICIS View 2014.1, ICIS View 2014.2

Distribution

US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Greece, Hong Kong, Iceland, Ireland, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Nicaragua, Norway, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Tanzania, Tunisia, United Arab Emirates, United Kingdom

States: Georgia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0055-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.