FDA Medical devices recall ·

Welch Allyn Life2000 Ventilation System recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-0046-2025
Date
(initiated September 12, 2024)
Company / brand
Baxter Healthcare Corporation
1 Baxter Pkwy, Deerfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
10,957 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP

Why was it recalled?

The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI/DI 00887761978089 or 00815410020537, All Life2000 Ventilators with software version 06.08.00.00

Distribution

US nationwide distribution.

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-0046-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.