FDA Medical devices recall ·

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10… recall

Recall number
Z-0026-2025
Date
(initiated September 12, 2024)
Company / brand
Becton Dickinson & Co.
7 Loveton Cir, Sparks, MD, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
162 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

BD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standalone laboratory software solution providing data management and workflow management functionality across locations where clinical diagnostic activities take place. Catalog Number: 444150

Why was it recalled?

Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00382904441500 Software versions: 4.20, 4.30, 5.10, 5.20 and 5.30

Distribution

Nationwide Foreign: Australia, New Zealand, Singapore, Hong Kong, Taiwan, Japan, Austria, Belgium, Denmark, France, Germany, Italy, Luxembourg, Netherlands, Norway, Spain, Sweden, Switzerland, United Kingdom, Russia, Saudi Arabia, South Africa, Canada, Mexico, China

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0026-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.