FDA Medical devices recall ·
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24JP recall
- Recall number
- Z-0016-2025
- Date
- (initiated June 26, 2024)
- Company / brand
- Smiths Medical ASD, Inc.
6000 Nathan Ln N, Minneapolis, MN, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 8976 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24JP
Why was it recalled?
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI/DI NA, Lot/Serial Numbers: 4013366, 4034019, 4037749, 4034021, 4040248, 4042841, 4048891, 4052407
Distribution
Worldwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.