FDA Medical devices recall ·

Allura Xper CV20 recall

Recall number
Z-0014-2026
Date
(initiated September 3, 2025)
Company / brand
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, Best, N/A, Netherlands
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
53 units (OUS only)
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Allura Xper CV20; Model Numbers: 722031; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Why was it recalled?

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model Numbers: 722031; UDI-DI: N/A; Serial Numbers: 29, 77, 23, 18, 45, 90, 164, 30, 76, 32, 75, 33, 40, 13, 113, 111, 46, 25, 130, 26, 61, 78, 42, 167, 49, 64, 41, 71, 20, 17, 63, 39, 9, 55, 36, 73, 48, 53, 31, 96, 28, 154, 166, 15, 84, 47, 35, 24, 148, 50, 79, 62, 66;

Distribution

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.

States: Georgia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.